New Aspiret suction pump technical guide
The films and troubleshooting guide below are designed to help you fix issues with the New Aspiret suction pump. You’ll find demonstrations on diagnosing faults, making repairs and installing replacement parts.
This suction pump is used to clear the nostrils and airways of babies, particularly when they are on respiratory support with oxygen or CPAP. It is an essential device that saves lives on neonatal units in hospitals.
The guides enable you to keep the New Aspiret suction pump in safe working order.
Check Warranty Before Repairing: For devices under warranty or service contract, unauthorized repairs may void coverage. Contact the manufacturer or authorized service provider before repairing.

Index of New Aspiret suction pump technical films
The following set of technical films offer instructions on how to access, check, repair and replace various components of the machine.
Troubleshooting guide for the New Aspiret suction pump
If you can find the issue you’re experiencing from the following suggestions, you can click to be directed to the most relevant video to help you solve the problem. You might want to watch the Overview film first to understand how the device works.
The suction pump does not power on
First, check that the machine is plugged in properly and switched on at the wall.
Next, watch ‘Checking the electrical circuit and replacing the power switch‘
The machine turns on but there is little or no suction
First, check that all tubing is firmly connected and free from leaks, kinks or blockages.
The pump is leaking liquid or there are signs of liquid damage
Stop using the device immediately and disconnect it from the power supply.
Watch ‘Cleaning the device after liquid damage‘
Next, watch ‘Replacing the diaphragm and the gasket‘
Next, watch ‘Disassembling the rest of the pump and cleaning the internal parts’